Designed to meet the requirements of
• HTM 02-01: Medical Gas Pipeline Systems
• C11 NHS: Model Engineering Specification – Medical Gases
• BS EN 6834: Specification for Active Anaesthetic Gas Scavenging Systems
• ISO 7396-2: Anaesthetic Gas Scavenging Disposal Systems
| Product Code | Plant Size |
|---|---|
| AGSS-*S/D*-130 | 130 l/min |
| AGSS-*S/D*-520 | 520 l/min |
| AGSS-*S/D*-1050 | 1050 l/min |
| AGSS-*S/D*-1500 | 1500 l/min |
| AGSS-*S/D*-1820 | 1820 l/min |
| AGSS-*S/D*-2500 | 2500 l/min |
| AGSS-*S/D*-3770 | 3770 l/min |
| AGSS-S/D-REMOTE | AGSS Remote Switch |
| *S/D* | S = Simplex / D = Duplex |
| Example: AGSS-D-520 | Duplex AGSS – 520l/min |
Specific Benefits Include
• CPX® branded solid British design
• Control and highly visible indication panel
• CE marked under Medical Device Directive 93/42/EEC with approval from British Standards Institute CE0086 classified as Class IIa Medical Device
• All interconnecting pipework is included
• Copper stubbed outlet and inlet pipework for ease of connecting to MGPS
• Available in Single or 3 Phase
• Anti-vibration mounts on all pumps
• Isolator to each pump unit as standard
• Alarm and BMS outputs for remote status indication
• Pre-set vacuum relief valve and low pressure sensors
• Low voltage (24V) remote switches
• Visible condensate drain flask fitted to exhaust
• Easy to install 4 wire + earth remote switch wiring
• All settings are factory set and verified with certificates of inspection
• Each plant has a full warranty of 12 months from its commission date